Applications:
· Validation logger for Measurement in the steam sterilizer
| Availability: | |
|---|---|
Product Description

Validation logger for thermal testing of processes with steam sterilizers. The logger is characterized by its highly flexible cable sensors, which enable precise positioning. With its 3 temperature sensors and the pressure sensor it is ideally suited for measurement in small sterilizers. The very thin temperature sensors with a diameter of 1.2 mm at the sensor tip enable a very fast response time. When using the logger in the washer-disinfector, it is strongly recommended to fix the sensors by means of silicone cable ties.
Benefits:
· highly flexible cable probes
· high accuracy
· Realtime data transmission
· fast response time
· sterilizable data logger
Applications:
· Validation logger for Measurement in the steam sterilizer
Data memory
4 x 20.064 measurement values
Sensor
Pt 1,000, class A
Weight
approx. 120 g
Accuracy
±0,1 °C (0 °C ... +120 °C) ±0,05 °C (+120 °C ... +140 °C) ±0,1 °C (+140 °C ... +150 °C)
±15 mbar (1 mbar ... 50 mbar), ±10 mbar (50 mbar … 150 mbar), ±10 mbar (2.050 mbar … 2.250 mbar),
±10 mbar (3.000 mbar … 3.250 mbar), ± 15 mbar (3.250 mbar ... 3.500 mbar),
± 0,5 % FS (3.500 mbar ... 4.000 mbar)
Protection class
IP 68
Dimensions
(Ø x H) 48 mm x 32 mm - Needle length 600 mm
Resolution
0,01 °C 1 mbar
Certificate
Factory Calibration: 0 °C, +60 °C and +134 °C 100 mbar and 3,100 mbar at +25 °C and 3,100 mbar at +134 °C
Measured variables
Temperature, pressure
Intervall
250 ms ... 24 h
Battery
3.6 V lithium battery exchangeable
Measuring range
Temperature 0 °C … +150 °C, pressure 1 … 4.000 mbar
Storage temperature
-20 °C ... +70 °C